beautypg.com

A qua m etrix, Ultrason page 19, Flange mount – Metex 3200U Ultrasonic Flow & Level Monitor User Manual

Page 19: Cross bar mount, Angle mount, Conduit mount, Stand pipe mount

background image

AquaMetrix

Ultrason

Page 19

4" OR 6" BLIND FLANGE

TAPPED ¾" NPT

FLEXIBLE
CONDUIT

JUNCTION

BOX

(OPTION JB)

ISOLATION
COUPLING
(SUPPLIED)
MUST BE
USED

DO NOT

CLAMP IN

THIS AREA

¾" NPT

NIPPLE

FLANGE MOUNT

ISOLATION
COUPLING
(SUPPLIED)
MUST BE
USED

DO NOT

CLAMP IN

THIS AREA

CLAMP

3/4"
CONDUIT

CROSS BAR MOUNT

ISOLATION
COUPLING
(SUPPLIED)
MUST BE
USED

DO NOT

CLAMP IN

THIS AREA

3/4" NPT

NIPPLE

ANGLE MOUNT

ISOLATION
COUPLING
(SUPPLIED)
MUST BE
USED

DO NOT

CLAMP IN

THIS AREA

3/4"
CONDUIT

CONDUIT MOUNT

Notes:

1. Use the ¾" NPT "Isolation Coupling"

supplied and

only. Do

not clamp sensor body or stem.

2. Locate the sensor 1 ft (30 cm) from

the sidewall or obstruction for every
10 ft (3 m) depth.

3.

mount in direct sunlight.

4. Extend sensor cable up to 500 ft

(150 m) with RG62AU coaxial only.

hand tighten

Do not

GRIND OR FILE

PIPE EDGE

SMOOTH

STAND PIPE MOUNT

STANDPIPE LENGTH
AS SHORT AS POSSIBLE
STANDPIPE DIAMETER
AS LARGE AS POSSIBLE

TYPICAL STANDPIPE:
4” / 100 mm DIAMETER
12” / 300 mm LENGTH

NARROW DIAMETER
STANDPIPES (<4” / 100 mm)
MAY AFFECT ACCURACY