QUICK BENCH TEST:

Connect Sensor to the

T.DUCER

terminals as shown below, then apply Power. When properly connected

a soft clicking can be heard from the sensor and figures will show on the large LCD display. Test
operation of the Ultrason by holding the sensor steadily and aiming at a flat, stable target 12 to 28" (305
to 711 mm) away from the end of the sensor. Allow a few seconds for the Ultrason to lock onto the
target before displaying its distance. The Ultrason will now display distance in ft or cm (factory
calibration).

CONNECTIONS:

POWER INPUT: The standard model requires AC power input between 100-130VAC 50/60Hz (2 amp
fuse is recommended). No adjustments are necessary for any voltage in this range. Optional 230VAC
requires power input between 200-260VAC 50/60Hz. (See OPTIONS section of this manual for
connection of optional 12VDC or 24VDC power input).

IMPORTANT NOTE: To comply with CSA/NRTL standards, AC power input and relay connection
wires must have conduit entry to the instrument enclosure.

AquaMetrix

Ultrason

Page 3

T.DUCER

N.O C N.C N.O C N.C N.O C N.C

RELAY 1

RELAY 2

RELAY 3

L N G

AC

RX PU GND TX

RS 232

N.O C N.C N.O C N.C N.O C N.C

RELAY 4

RELAY 5

RELAY 6

4\20

GROUND STUD

:

ATTACH CABLE
SHIELDS (4-20mA, RS232)
THIS END ONLY

IMPORTANT

IMPORTANT:
MUST CONNECT TO A
GOOD GROUND (<1 Ohm)
WITH 12 AWG CONDUCTOR

OPTIONAL

100-130VAC (50/60 Hz)

OPTIONAL:
200-260VAC (50/60 Hz)

*

*

A

M

qua etrix

ALUMINUM

CHASSIS