CALIBRATION
- for Level/Inventory Applications
1. Before starting the calibration determine:
a)
MAX RANGE = ________________
(Maximum range = distance from Sensor to Zero
level)
b)
MAX LEVEL
= ________________
(Maximum level of product being measured)
c)
MIN RANGE
= ________________
(Distance from sensor to Max Level)
Minimum range = MAX RANGE - MAX LEVEL
(must be at least 8" / 203 mm depending on sensor
model)
2. Check the maximum range with the sensor installed by:
a)
Use
Æ to get to
UNITS/MODE
then
È to get to
Range
, then
Store?
Æ
Yes
(NOTE: liquid must be at zero level)
or
b)
Carefully measure distance from sensor to zero
level with tape measure.
MINIMUM RANGE:
At
CALIBRATION
- Press
È to get to
MinRg
- Press
Æ to move the cursor under the digits
- Use
È or Ç to set the minimum range
Note:
MinRg
must be
³12" (30.5 cm) for US2-1A1A
sensors and
³ 16" (40.6 cm) for US2-2A1B sensors except
in
Hrt
mode. For correct tank volume calculation in
Hrt
mode
MinRg
must be the actual distance from the end of
the sensor to the top of the tank (any value
³0). In
Level
or
Range
modes the Ultrason will only accept
MinRg
values
³ 8" (20.3 cm).
- Press
Æ to return to
MinRg
MAXIMUM RANGE:
Press
È to get to
MaxRg
Repeat as for
MinRg
AquaMetrix
Ultrason
Page 10
MAX LEVEL
ZERO LEVEL
MIN RANGE
TANK INVENTORY
RANGE MODE
SENSOR
0%
100%
MAX RANGE
MAX LEVEL
ZERO LEVEL
MAX
VOLUME
MAX RANGE
MIN RANGE
TANK INVENTORY
LEVEL MODE
SENSOR
100%
0%
TANK INVENTORY
HORIZONTAL ROUND TANK (HRT) MODE
MAX RANGE
MAX
VOLUME
MIN RANGE