TECHNICAL DATA

7

57

OPTISYS TUR 1050

www.krohne.com

11/2012 - 4002444301 MA OPTISYS TUR 1050 R02 en

Operating conditions

Note: the device is not appropriate for outdoor use!

Temperature

Process and ambient

temperature

+1...+50°C / +34...+122°F

Storage temperature

-20...+60°C / -4...+140°F

Pressure

Process pressure

0.07...14 bar / 1...200 psi (built in regulator set at 1 bar / 15 psi)

Ambient pressure

Atmospheric, altitude up to 2000 m / 6600 ft

Flow velocity

Minimum

0.1 l/min / 0.026 gal/min

Maximum

1 l/min / 0.26 gal/min

Other conditions

Humidity

Display and operating unit: up to 95% relative humidity (non-condensing)

Protection category

Enclosure is designed to meet IP 66 / NEMA 4X

Installation conditions

Dimensions

For detailed information refer to

Dimensions and weight

on page 58.

Weights

Shipping weight: 2.5 kg / 5.5 lbs

Materials

Sensor housing

ABS

Wetted parts

Nylon, borosilicate glass, silicon, polypropylene, stainless steel AISI 304

Electrical connections

Galvanic insulation

Double insulated, pollution degree 2, over voltage category II (all in- and outputs

are electrically isolated from each other and from power and earth ground)

Option: transformer isolated current output

Power supply

Voltage

100...240 VAC, 47...63 Hz

Power consumption

80 VA

Outputs

Current output

1 x 4...20 mA, active, 15 VDC power source, max. load 600 Ω

Modbus

Bi-directional, RS-485 Modbus RTU/ASCII

Relays

2 x relay 120...240 VAC, freely programmable

Approvals and certifications

CE

This device fulfils the statutory requirements of the EC directives. The

manufacturer certifies successful testing of the product by applying the CE mark.

ETL

Listed to UL 61010B-1 and certified to CSA 22.2 No. 1010.1-92