Operation – KROHNE OPTISYS TUR 1050 EN User Manual

Page 45

background image

OPERATION

5

45

OPTISYS TUR 1050

www.krohne.com

11/2012 - 4002444301 MA OPTISYS TUR 1050 R02 en

In the communication mode "Simple" the following information is exchanged between the
master computer and the device:

Desiccant alarm

The device can output a humidity alarm if the detected humidity could cause condensation. The
default setting of the humidity alarm is "OFF", like in the following drawing:

If you enabled the desiccant alarm and the humidity detector determines a problematic value,
the device activates the alarms and sets the 4...20 mA output to the selected error level. In the
normal measuring mode you see the measured value in the upper row and "DESC" in the lower
row:

Enquiries of master computer

Answer of device

Byte 1: attention character ":" in ASCII or 3A Hex

The same attention character ":" in ASCII or 3A Hex

Byte 2: address of the device being enquired

Address of the device

Byte 3/4: CR/LF or 0D 0A in Hex

3: Measured value

4: Unit (NTU or FNU)

Example (master requesting a report from address

No. 1): 1 CRLF

Example (response of the device set to address

No. 1): 001 0.0249 NTU

AUTO

CAL

CONFIG