beautypg.com

Installation – KROHNE OPTISYS TUR 1050 EN User Manual

Page 21

background image

INSTALLATION

3

21

OPTISYS TUR 1050

www.krohne.com

11/2012 - 4002444301 MA OPTISYS TUR 1050 R02 en

3.4.6 Inserting and fixing the ultrasonic cuvette with flow-through holder

CAUTION!
Never insert a cuvette with visible moisture or water on the glass or the transducer into the
upper part of the device, because this could destroy or damage the electronics or the
transducer! Always clean and dry the cuvette with a soft cloth just before inserting it. The
dehumidification system cannot remove large droplets of water, only residual moisture!

Figure 3-5: Procedure for inserting and fixing the ultrasonic cuvette with flow-through holder

INFORMATION!
The device can only detect a new cuvette if it works in the normal operation mode ("AUTO"). If the
device operates correctly after inserting a new cuvette, "AUTO" will flash on the display.