beautypg.com

2 technical data table, Technical data – KROHNE OPTISYS TUR 1050 EN User Manual

Page 56

background image

7

TECHNICAL DATA

56

OPTISYS TUR 1050

www.krohne.com

11/2012 - 4002444301 MA OPTISYS TUR 1050 R02 en

7.2 Technical data table

INFORMATION!

The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.

Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).

Measuring system

Measuring principle

90° scattered light method

Application range

Turbidity measurement of liquids according to ISO 7027 and US EPA 180

Measuring range

0.02...100 NTU/FNU or 0.02...1000 NTU/FNU

Design

The measurement system consists of a sensor and a signal converter and is only available as compact version.

Display and user interface

Display

Multi-line liquid crystal backlight display

Operating and display language

English

Other properties

Response time

Min. 5 seconds, adjustable

Alarms

Two programmable, 120...240 VAC, 2A form C relay

Measuring accuracy

Reference conditions

+1...+50°C / +33.8...+122°F

95% relative humidity

Altitude up to 2000 m / 6600 ft

Accuracy

< 40 NTU/FNU: ±2% of reading or ±0.02 NTU/FNU whichever is greater

> 40 NTU/FNU: ±5% of reading

Resolution

0.0001 NTU/FNU (below 10 NTU/FNU) selectable