2 technical data table, Technical data – KROHNE OPTISYS TUR 1050 EN User Manual
Page 56
7
TECHNICAL DATA
56
OPTISYS TUR 1050
www.krohne.com
11/2012 - 4002444301 MA OPTISYS TUR 1050 R02 en
7.2 Technical data table
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Measuring principle
90° scattered light method
Application range
Turbidity measurement of liquids according to ISO 7027 and US EPA 180
Measuring range
0.02...100 NTU/FNU or 0.02...1000 NTU/FNU
Design
The measurement system consists of a sensor and a signal converter and is only available as compact version.
Display and user interface
Display
Multi-line liquid crystal backlight display
Operating and display language
English
Other properties
Response time
Min. 5 seconds, adjustable
Alarms
Two programmable, 120...240 VAC, 2A form C relay
Measuring accuracy
Reference conditions
+1...+50°C / +33.8...+122°F
95% relative humidity
Altitude up to 2000 m / 6600 ft
Accuracy
< 40 NTU/FNU: ±2% of reading or ±0.02 NTU/FNU whichever is greater
> 40 NTU/FNU: ±5% of reading
Resolution
0.0001 NTU/FNU (below 10 NTU/FNU) selectable