5.7.3 Configuring the RS 485 port
If you selected the RS 485 output (for detailed information refer to
Selecting the output
on page
36), the device automatically switches to this submenu. It allows to configure the RS 485 output
and starts with the baud rate:
• Use the buttons ↑ and ↓ to select one of the predefined baud rates or leave the setting
(options: 1200, 2400, 4800, 9600 or 19200 baud).
• Accept your selection by pressing ^.
i
The device switches to the next menu item, i.e. the setting of the device address.
• Use the buttons ↑ and ↓ to select the desired device address or leave the setting (options:
1...255).
• Accept your selection by pressing ^.
i
The device switches to the last menu item of this submenu, i.e. "Modbus mode".
• Use the buttons ↑ and ↓ to select the options "RTU" or "ASCII" or leave the setting.
• Accept your selection by pressing ^.
i
The device switches to the submenu "Configuring the alarm relays".
INFORMATION!
The default communication parameters of the RS 485 port are: 8 bits, no parity and 1 stop bit.
AUTO
CAL
CONFIG
INFORMATION!
For further information concerning the Modbus mode contact the manufacturer.