1 notes on installation, 2 storage and transport, 3 requirements of the device – KROHNE OPTISYS TUR 1050 EN User Manual
Page 14: Installation
3
INSTALLATION
14
OPTISYS TUR 1050
www.krohne.com
11/2012 - 4002444301 MA OPTISYS TUR 1050 R02 en
3.1 Notes on installation
3.2 Storage and transport
• Store and transport the device in a dry, dust-free environment.
• Store and transport the device in an environment with a temperature between -20...+60°C /
-4...+176°F.
• The original packing is designed to protect the equipment. It has to be used if the device is
transported or sent back to the manufacturer.
3.3 Requirements of the device
The device was designed to require very low input pressure. Though the device offers a wide
pressure range, as it has an integrated pressure regulator:
• Input pressure range: 0.07...14 bar / 1...200 psi (built in regulator set at 1 bar / 15 psi)
• Maximum allowable flow rate of the cuvette: 0.1...1 l/min / 0.026...0.26 gal/min
• Maximum fluid temperature: +50°C / +122°F
INFORMATION!
Inspect the cartons carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.