6. Troubleshooting

39

6. Troubleshooting

6.1 Check Power

Symptom

Possible Cause / Solution

Refer to

- When using battery: The battery is exhausted.
=> Turn the power off and charge the battery.

4

Even when the power switch
is turned ON, on-screen
display does not appear or
display goes off quickly.

- When using AC adapter: The dedicated adapter is not being used.
=> Make sure to use the dedicated AC adapter.

5

Even after replacing batteries
with new ones, the screen
does not show up.

- The polarity of the batteries may be reversed.
=> Check if the batteries are inserted properly.

4

Batteries die very quickly.

Different type of batteries may be used.
Exhausted batteries and new batteries may be mixed.

4

The batteries may not have been charged fully.
=> Charge the batteries fully referring to the operation manual supplied
with the NiMH batteries.

4

When using NiMH batteries,
the batteries die very quickly.

The NiMH batteries may be deteriorated.
=> Purchase new NiMH batteries.

4

6.2 Check while Measuring

Symptom

Possible Cause / Solution

Refer to

Rubber cap may be put on the inlet.
Remove the rubber cap and then insert the removed rubber cap to the
cleaning air vent to perform a measurement.

2

Count value remains 0.

The exhaust port may be blocked.
Remove the things that block the exhaust port.

2

The rubber cap may not be inserted to the cleaning air vent.
Insert the rubber cap properly before starting a measurement.

2

Count value is too low.

Compare the count value upon checking the sensitivity to the standard
count value.
If the count value upon checking the sensitivity is more than 5% lower
than the standard value, calibrate the unit to check the sensitivity.

8

The sensitivity check knob may still be at “FULL” side.
Turn the knob fully to the “ZERO” side.

2

Count value is too high.

Compare the count value upon checking the sensitivity to the standard
count value.
If the count value upon checking the sensitivity is more than 5% higher
than the standard value, calibrate the unit to check the sensitivity.

8