3. Display Description and operation Procedure

31

F) Pulse Output

With this instrument the pulse that corresponds to CPM value (instantaneous value) can be output.

Connect the output cable sold separately.

(1) Pulse Frequency Update Interval: The output data is updated every one second.
(2) Pulse Range:

0.0166 Hz – 166.6 Hz (1-10,000 CPM)
DUTY 50%

(3) Output Format, Spec:

Photocoupler Output (no-voltage contact output: collector emitter)

Model: TLP127
Maximum Applied Voltage: DC300V
Maximum Current: 150mA

(4) Output

Cable:

Please

purchase

our

optional

output

cable

(3442-01).

Pulse Output

Explanation

CPM (Instantaneous Value)

Pulse Frequency

1 CPM

0.0166 Hz

60 CPM

1 Hz

600 CPM

10 Hz

6,000 CPM

100 Hz

10,000 CPM

166.6 Hz

* The pulse output is more than 1 CPM and the resolution is 1

CPM. When CPM is 0 CPM, there is no pulse output.


The pulse equivalent to the CPM
value (instantaneous value) is
output.

(Output Terminal Circuit)






G) Output Cable (3442-01) -external appearance-

This is the dedicated cable for analog output, pulse output and alarm output. This is an option at the time of
purchase only.










* The CPM value (instantaneous value) used for pulse output differs

from the COUNT value on the LCD display.

* Use the dedicated output cable (3442-01) sold separately.

output

1000 mm

150 mm

Cable 1

Cable 2

Relay Connector

Bare crimping terminal – Y type (forM3)

Connector at the
dust monitor side

* Supplied as a set of cable 1

and cable 2.

Signal Name

Wiring Color ‐

Cable 2

Marking

Sig Red ANALOG +

Analog Output

GND

Black

ANALOG −

White

ALARM +

Alarm Output

Blue

ALARM −

Yellow

PULSE +

Pulse Output

Green

PULSE −