3. Display Description and operation Procedure
29
(1) Data Update Interval: The output data is output every one second.
(2) Load Impedance:
1 KΩ or higher
(3) Output Cable:
Please purchase our optional output cable (3442-01).
Method to output measurement data (in case of analog output)
Explanation
Concentration value
(mg/m
3
or CPM) will be
converted to the output
voltage.
When the reading exceeds the set concentration value during a
measurement, the alarm will be turned ON. And when the reading
becomes lower than the set concentration value, the alarm will be OFF.
This alarm will be activated only during a measurement.
(Even if the alarm is ON during a measurement, it will be OFF once the
measurement finishes.)
When alarm mode setting is OFF, the alarm function is not activated.
Use [▲] [▼] keys to change the mode setting.
After setting the alarm mode, press [SET] key.
The RANGE setting value will be highlighted. Use [▲] [▼] keys to set
the value. After setting, press [SET] key. Then the RANGE setting value
will stop being highlighted.
0 1 2 3 4 sec(Measurement Time)
Voltage
* Use the dedicated output cable (3442-01) sold separately.
mg/m3
CPM
OFF
[▲] key to CPM setting ↓ ↑ [▼] key to mg/m3setting
[▲] key to OFF setting ↓ ↑ [▼] key to CPM setting
10.000
9.999
[▼] key to decrease ↓ ↑ [▲] key to increase