Temperature ranges
1
Operational: -20-80° C (-4-176° F)
Storage: -40-80° C (-40-176° F)
Calibrated: -5-50° C (23-122° F)
Diameter
1.83 cm (0.72 in.)
Length
21.6 cm (8.5 in.)
Weight
197 g (0.43 lb)
Materials
Titanium body; Delrin nose cone
Output options
Modbus/RS485, SDI-12, 4-20 mA
Battery type & life
2
3.6 V lithium; 10 years or 2M readings
External power
8-36 VDC
Memory
Data records
3
Data logs
2.0 MB
130,000
50
Log types
Linear, Fast Linear, and Event
Fastest logging rate & Modbus rate
2 per second
Fastest SDI-12 & 4-20 mA output rate 1 per second
Real-time clock
Accurate to 1 second/24-hr period
Sensor Type/Material
Piezoresistive; titanium
Calibrated Range
(Usable Depth)
Gauged (vented)
5 psig (3.5 m, 11.5 ft)
15 psig (11 m, 35 ft)
30 psig (21 m, 69 ft)
100 psig (70 m, 231 ft)
300 psig (210 m, 692 ft)
500 psig (351 m, 1153 ft)
Burst pressure
Max. 2x range; burst > 3x range
Accuracy @ 15° C
4
±0.1% FS
Accuracy (FS)
5
±0.05% FS
Resolution
±0.005% FS or better
Units of measure
Pressure: psi, kPa, bar, mbar, mmHg, inHg, cmH
2
O, inH
2
O
Level: in, ft, mm, cm, m
Temperature Sensor
Accuracy & resolution
±0.1° C; 0.01° C or better
Units of measure
Celsius or Fahrenheit
Warranty
3 years
Up to 5-year extended warranties are available for all instruments—call for details
Footnotes