Range
Usable Depth
PSIG
kPA
Meters
Feet
5
34.5
3.5
11.5
15
103.4
11
35
30
206.8
21
69
100
689.5
70
231
300
2068
210
692
500
3447
351
1153
Vented Level TROLL Instrument
Verify Instrument Depth
After you have installed the instrument, it is possible to connect the instrument to a
computer or RuggedReader, open the software, and take a reading to verify the
installation position. If the reading confirms that the instrument is in the correct position,
you can secure it as described below.
During log setup there was an option to "Remind Me Later" for setting a level reference.
If you set the log to remind you later, ensure that the instrument is submerged and set
the level reference when prompted. See page 53.
Secure the Cable
The RuggedCable System includes a Kellems grip near the up-hole end. You can
compress the wire mesh and slide the grip to the desired location on the cable. Pull on
the grip to tighten it on the cable.
Use the loop on the Kellems grip to anchor the cable to a convenient stationary object
such as the In-Situ well dock installation ring. To install the Kellems grip to the
installation ring, insert the loop into the locking clip on the well dock and position the
assembly at the top of a well casing.