In-Situ Level TROLL 400/500/700/700H Operators Manual User Manual
Page 78
1-800-446-7488
78
www.in-situ.com
Problem
Possible Cause
Possible Solution
Cannot configure level or other
parameters using the Configure
button on the Sensors tab. The
Sensor Setup screen is shown, but
the Configure... button is dim.
The instrument is actively polling
(continually updating real-time readings)
in the Home tab.
The instrument has an active log—a log
that is Ready, Pending, Running, or
Suspended as shown in the Status
column of the Logging tab. Only one
active log can reside in the device at a
time.
Return to the Home tab and stop real-time
readings by clicking the Play button.
Stop or delete the log if possible. Alternatively,
configure parameters after the log is complete.