6 field deployment, Program the instrument, Position the instrument – In-Situ Level TROLL 400/500/700/700H Operators Manual User Manual
Page 34: Field deployment
1-800-446-7488
34
www.in-situ.com
Field Deployment
Program the Instrument
In order to set up a log or download data, you must connect the instrument to a
computer running Win-Situ 5 Software, or a RuggedReader Handheld PC running Win-
Situ Mobile. See page 50.
Position the Instrument
Place the instrument at the desired depth. Position the instrument below the lowest
anticipated water level, but not so low that the pressure sensor range might be
exceeded at the highest anticipated water level. Refer to the tables below to determine
usable depth.
A BaroTROLL Instrument can be deployed with a non-vented
instrument to compensate level data for changes in atmospheric
pressure. Make sure the clocks on both instruments are
synchronized, and install the BaroTROLL in a location that will
never be submerged in water.
Range
Effective Range
Usable Depth
PSIA
PSIA
kPA
Meters
Feet
30
15.5
106.9
11
35
100
85.5
589.5
60
197
300
285.5
1968
200
658
500
485.5
3347
341
1120
1000
985.5
6795
693
2273
* Effective range for psia sensors is limited by an estimated
14.5 PSI atmospheric pressure at sea level.
Non-Vented Level TROLL Instrument