beautypg.com

GW Instek GDM-8246 Programming Manual User Manual

Page 13

background image

GDM-8246 MULTIMETER

PROGRAMMER MANUAL

⎯ 24 ⎯

*SRE?

Returns:

is in the range from 0 through 255.

Examples

*SRE 7 sets bits of the SRER to 0000 0111.

If the *SRE? returns 3, the SRER is set to 0000 0011.

*STB?

(query only)

Function:
The query of the Status Byte register (SBR) with *STB? will return a
decimal number representing the bits that are set (true) in the status
register.

Syntax:

*STB?

Returns:

is in the range from 0 through 255.

Examples:

*STB? returns 81, if SBR contains the binary value 0101 0001.

*WAI

(no query form)

Function:
WAI prevents the programming instrument from executing further
commands or queries until all pending operations are finished.

Syntax:

*WAI

GDM-8246 MULTIMETER

PROGRAMMER MANUAL

⎯ 25 ⎯

:CONFigure:AUTo

Function:
Set or query the auto-range mode on or off.

Syntax:

:CONFigure:AUTo?

:CONFigure:AUTo

can be 0(off) or 1(on)

Returns:

0/1

Examples:

:CONFigure:AUTo 1

Set to auto-range mode.

If it is in the auto-range mode, the command of :CONFigure:AUTo? Will
return the value of 1.

:CONFigure:RANGe?(query only)

Function:
Set or query the range of the present function.

Syntax:

:CONFigure:RANGe?

Returns: