GDM-8246 MULTIMETER

PROGRAMMER MANUAL

⎯ 20 ⎯

z Miscellaneous Commands

Table 4 lists the miscellaneous commands that control general
housekeeping functions of the Digital Multimeter.

Table 4: Miscellaneous Commands

*IDN?

Returns instrument identification.

*OPC

Reports when operation is completed by
setting the Operation Complete bit in SESR.

*OPC?

Reports when operation is completed. Same as
*OPC except returns a 1 to the output queue
and dose not set the SESR bit.

*RST

Resets the protection levels and states, resets
the current and voltage levels to zero, sets the
output off, and sets memory section to 00.

*WAI

Wait to continue. This command forces
sequential operation of commands. This
command is required by IEEE-488.1-1987.
The DMM, however, forces sequential
operation of commands by design.

:SYSTem:ERRor?

Read the next item from the error/event queue.

:SYSTem:VERSion?

Returns the SCPI version level.


GDM-8246 MULTIMETER

PROGRAMMER MANUAL

⎯ 21 ⎯

6. DETAILS OF COMMAND REFERENCE

Each command in this chapter will give a detailed description. The

examples of each command will be provided and what query form might
return.

*CLS

(no query form)

Function:
Clear all event status data register. This includes the Output Queue,
Operation Event Status Register, Questionable Event Status Register, and
Standard Event Status Register.

Syntax:

*CLS

Examples:

*CLS clears all event registers.

*ESE

Function:
Set or return the bits in the Event Status Enable Register (ESER). The
ESER enables the Standard Event Status Register (SESR) to be
summarized on bit 5 (ESB) of the Status Byte Register (SBR).

Syntax:

*ESE

*ESE?

is in the range from 0 through 255.

Returns:

is a number from 0 to 255 that indicates the decimal value of the
binary bits of the ESER.