GW Instek GPI-700A Series Programming Manual User Manual
Page 8
GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
⎯ 13 ⎯
Combining Commands
You can use a semicolon (;) to combine commands. But continuously
query command will cause message missing. For example
:GRO:STEP: ACWS:VOLT:;CMAX?
If the command that follows the semicolon has a different header path
from the root level, you must use a colon to force a return to the root
level.
:ACC:CMIN 1.00;:ACC:CMAX 2.00
If the command that follows the semicolon has the same header path, you
may omit the colon and the path and state only the new leaf node. For
example:
:ACC:CMIN 1.00;:ACC:CMAX 2.00
is equal to
:ACC:CMIN 1.00;:CMAX 2.00
You can combine commands and queries into the same message. Note,
for example, the following combination:
:ACC:CMIN 1.00;CMIN?
Synopsis of Commands
The tables in this section summarize the command of the programmable
power supply. These tables divide the commands into three functional
classifications:
z
General Setting Commands
z
Status Commands
z
Miscellaneous Commands
The tables also provide a brief explanation of each command.
GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
⎯ 14 ⎯
z General Setting Commands
Table 2 lists the general setting commands that control and query the
settings of the power supply.
Table 2: General Setting Commands
Commands
Explanation
:GROup
Set the value of group
:GROup?
Query the value of group
:STEP
Set the value of step
:STEP?
Query the value of step
:FUNCtion:TEST:MODe
Set the value of test mode
:FUNCtion:TEST:MODe?
Query the value of test mode
:FUNCtion:TEST:STATe
Set the value of test state
:FUNCtion:TEST:STATe?
Query the value of test state
:MEASure
Query the value of measurement
:UTILity: FREQuency <50 or 60>
Set the value of AC frequency
:UTILity: FREQuency?
Query the value of AC frequency
:UTILity:ARC:CURRent
Set the value of ARC current
:UTILity:ARC:CURRent?
Query the value of ARC current
:UTILity:ARC:MODe
Set the ARC test mode
:UTILity:ARC:MODe?
Query the ARC test mode
:UTILity:TEST:MODe
Set the test mode from step 1 or form
present step
:UTILity:TEST:MODe?
Query the test mode
:UTILity:IRTest:MODe
Set the value of IR test mode
:UTILity:IRTest:MODe?
Query the value of IR test mode