GPI/GPT SERIES PROGRAMMABLE EST

PROGRAMMER MANUAL

⎯ 29 ⎯

V/I:

Test Function

Unit (2 bytes)

ACW

KV

DCW

KV

IR

V

Cnt (GB)

A

CAC

KV

CDC

KV

HOLD

PAUSE

Meas:

Test Function

Unit (4 bytes)

ACW

mA

DCW

mA

IR

MOhm

Cnt (GB)

mOhm

CAC

mA

CDC

mA

HOLD

GPI/GPT SERIES PROGRAMMABLE EST

PROGRAMMER MANUAL

⎯ 30 ⎯

Time:

Time (1 byte)

Value (5 bytes)

Unit (1 bytes)

Ramp

005.0

sec

Test

005.0

sec

Examples:

User the command :MEAS3? after measurement

Display for PC 10,1,0.100kV,01.00mA, T005.0S

10=Pass

1=ACW

0.100kV=Mesuremett voltage 100V

01.00mA=Mesurement current 1mA

T005.0S=Test time 5 sec

:UTILity:AC:FREQuency

Function:

Set or query the ACW CAC frequency ( 0:50Hz, 1:60Hz )

Syntax:

:UTILity: FREQuency

:UTILity: FREQuency?

50 (50Hz) or 60 (60Hz)

Return:

50 | 60