GPI/GPT SERIES PROGRAMMABLE EST

PROGRAMMER MANUAL

⎯ 49 ⎯

Syntax:

:GBONd:RMINimum

:GBONd:RMINimum?

Please refer to the specification

Return:

Examples:

:GBONd:RMINimum 100.0 set the GB test maximum resistance to 100.0mΩ.

:GBONd:RMINimum? return 100.0 if the GB test maximum resistance setting is

100mΩ.

:GBONd:TTIMe

Function:

Set or query the GB test time.

Syntax:

:GBONd:TTIMe

:GBONd:TTIMe?

Please refer to the specification

Return:

Examples:

:GBONd:TTIMe 5.0 set the IR test time to 5 sec.

:GBONd:TTIMe? return 005.0 if the GB test time setting is 5 sec.

GPI/GPT SERIES PROGRAMMABLE EST

PROGRAMMER MANUAL

⎯ 50 ⎯

:SYSTem:ERRor?(Querly only)

Function:

Query next error message from the Error/Event queue. The result of the query is

the error number followed by the error text.

Syntax:

:SYSTem:ERRor?

Return:

Examples:

System:ERRor? Return 0, “No error”

:SYSTem:VERSion?(Querly only)

Function:

Return the SCPI version of the device.

Syntax:

:SYSTem:VERSion?

Return:

1994.0