GW Instek GPI-700A Series Programming Manual User Manual
Page 25

GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
⎯ 47 ⎯
:IRESistance:TTIMe
Function:
Set or query the IR test time.
Syntax:
:IRESistance:TTIMe
:IRESistance:TTIMe?
Return:
Examples:
:IRESistance:TTIMe 5.0 set the IR test time to 5 sec.
:IRESistance:TTIMe? return 005.0 if the IR test time setting is 5 sec.
:GBONd:CURRent
Function:
Set or query the GB test current.
Syntax:
:GBONd:CURRent
:GBONd:CURRent?
Return:
GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
⎯ 48 ⎯
Examples:
:GBONd:CURRent 25.00 set the GB test current to 25A.
:GBONd:CURRent? return 25.00 if the GB test current setting is 25A.
:GBONd:RMINimum
Function:
Set or query the GB test minimum resistance.
Syntax:
:GBONd:RMINimum
:GBONd:RMINimum?
Return:
Examples:
:GBONd:RMINimum 100.0 set the GB test minimum resistance to 100mΩ.
:GBONd:RMINimum? return 100.0 if the GB test minimum resistance setting is
100mΩ.
:GBONd:RMAXimum
Function:
Set or query the GB test maximum resistance.
