GW Instek GPI-700A Series Programming Manual User Manual
Page 19

GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
⎯ 35 ⎯
:ACContinue:CMINimum
Function:
Set or query the CAC test minimum current.
Syntax:
:ACContinue:CMINimum
:ACContinue:CMINimum?
Return:
Examples:
:ACContinue:CMINimum 1.00 set the CAC test minimum current to 1.00mA.
:ACContinue:CMINimum? return 01.00 if the CAC test minimum current setting
is 1.00mA.
:ACContinue:CMAXimum
Function:
Set or query the CAC test maximum current.
Syntax:
:ACContinue:CMAXimum
:ACContinue:CMAXimum?
Return:
GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
⎯ 36 ⎯
Examples:
:ACContinue:CMAXimum 1.00 set the CAC test maximum current to 1.00mA.
:ACContinue:CMAXimum? return 01.00 if the CAC maximum current setting is
1.00mA
:ACContinue:RTIMe
Function:
Set or query the CAC ramp time.
Syntax:
:ACContinue:RTIMe
:ACContinue:RTIMe?
Return:
Examples:
:ACContinue:RTIMe 5.0 set the CAC ramp time to 5 sec.
:ACContinue:RTIMe? return 005.0 if the CAC ramp time setting is 5 sec.
:DCContinue:VOLTage
Function:
Set or query the CDC test voltage.
Syntax:
:DCContinue:VOLTage