:FUNCtion:TEST:MODe? return 3 if the test function setting is IR.

:FUNCtion:TEST:STATe

Function:

Set or query the test state (0. OFF 1. ON 2. CONTINUE). During testing, when

proceed HOLD function, the instrument will stop testing until a command of

“FUNC:TEST:STAT 2” has been received.

Syntax:

:FUNCtion:TEST:STATe

:FUNCtion:TEST:STATe?

is in the range from 0 through 1.

Return:

Examples:

:FUNCtion:TEST:STATe 1 start test.

:FUNCtion:TEST:STATe ? return 1 if the test state is 1.

:MEASure

Function:

Return measurement value

Syntax:

:MEASure?

is in the range from 1 through 16 for STEP.

GPI/GPT SERIES PROGRAMMABLE EST

PROGRAMMER MANUAL

⎯ 28 ⎯

: MEASure? Return measurement value of COM mode.

Return:

Status, Mode, V/I, Meas, Time

(2 bytes) (1 byte) (7 bytes) (8 bytes) (7 bytes)

Status:

00:Emtpy

10:Pass

20:Fail

21:Interlock

22:Low Voltage

23:Hi voltage

24:ARC

25:Measurement over limit maximum

26:Measurement below limit minimum

30:Stop

32:Hight Voltage

33:Heatsink is too hot

Mode:

0: Empty

1:ACW

2:DCW

3:IR

4:Cnt (GB for 745A)

5:CAC

6:CDC

7:HOLD