1600A-N1R

1600A-N1R Instruction Manual

Rev. 0.1

Page 26 of 38

Figure 33 S7-200 USB/PPI Cable

5.6.1 Touch Screen Upgrade

1. Disconnect blue DB9 cable from PLC Communications Module (Figure 34).
2. Using a Gender Changer, connect the USB/PPI cable to the DB9 Cable disconnected from the PLC

Communications Module, and connect the USB end of the USB/PPI cable to a PC or laptop.

Enclosure Lid

PLC

Communications

Module

DB9 Cable Connector

Figure 34 Display Back Panel

3. From the touch screen display, switch to transfer mode by selecting:

Program Menu Ź Screen Utilities Ź Transfer

4. Open ProSave software on PC or laptop:

Start Ź All Programs Ź Simatic Ź ProSave Ź ProSave

5. Click on the “General” tab.
6. Under "Device Type", select TP 177A 6”.
7. Under "Connection", select Serial (via USB-PPI cable) for connection.
8. Click on the “Restore” tab.
9. Select the firmware file named “Detcon_1600AN1P_HMI_vXX.psb”. (XX represent current version.)
10. Ensure that “Booting” check box is unchecked.
11. Press “Star Restore” to start the firmware upgrade.