Detcon 1600A-N1R User Manual
Page 30
1600A-N1R
1600A-N1R Instruction Manual
Rev. 0.1
Page 26 of 38
Figure 33 S7-200 USB/PPI Cable
5.6.1 Touch Screen Upgrade
1. Disconnect blue DB9 cable from PLC Communications Module (Figure 34).
2. Using a Gender Changer, connect the USB/PPI cable to the DB9 Cable disconnected from the PLC
Communications Module, and connect the USB end of the USB/PPI cable to a PC or laptop.
Enclosure Lid
PLC
Communications
Module
DB9 Cable Connector
Figure 34 Display Back Panel
3. From the touch screen display, switch to transfer mode by selecting:
Program Menu Ź Screen Utilities Ź Transfer
4. Open ProSave software on PC or laptop:
Start Ź All Programs Ź Simatic Ź ProSave Ź ProSave
5. Click on the “General” tab.
6. Under "Device Type", select TP 177A 6”.
7. Under "Connection", select Serial (via USB-PPI cable) for connection.
8. Click on the “Restore” tab.
9. Select the firmware file named “Detcon_1600AN1P_HMI_vXX.psb”. (XX represent current version.)
10. Ensure that “Booting” check box is unchecked.
11. Press “Star Restore” to start the firmware upgrade.