DeFelsko PosiTector 100 v.4.0 User Manual
Page 31

100 Series
Page 31
G
AGE
INITIALIZATION
This is a destructive action. It not only restarts the internal software program,
but it also erases all stored user settings or data and returns the Gage to a known
out-of-the-box condition.
This action:
•
performs a Gage RESET as described above
•
erases all readings in memory
•
erases all stored probe information
•
erases all user created or downloaded configurations and list boxes
•
turns off Memory, Statistics and Graphics modes
•
begins application and batch names at number 1
•
erases all applications and restores default
App1 (1 coat of “polymer” on
substrate “none”)
•
is handy when uncertain what previous changes have been made to the
Gage such as when beginning a new job or taking possession of the Gage
from a previous owner/operator
NOTE: For Gages with the Memory option, stored measurements and
downloaded user configurations/lists are deleted during this
operation. If the user wishes to keep these values, they should
first be downloaded by PosiSoft onto a computer. List box
contents can be re-loaded into the Gage after the
“Initialization” has been successfully performed.
There are two ways to INITIALIZE the Gage. The first is by selecting the
Admin q
q Initialize Gage menu item. The second method is to first press
the RESET button and then power-up the Gage by simultaneously pressing and
holding the left-most AND right-most buttons until a BEEP is heard.