DeFelsko PosiTector 6000 v.4.0 User Manual
Page 9

9
How to Calibrate
Make several measurements on each of 
three to five traceable Calibration 
Standards (see below). 
If all of the readings on each standard is 
within the tolerance of the Gage, the 
Gage is operating properly and the 
calibration is complete. 
If the readings are not within tolerance, 
first verify that your standards are good, 
then repeat the measurements. If the 
measurements are still out of tolerance, 
you may wish to try a second set of 
standards. If after this step the Gage is 
still reading out of tolerance, the Gage 
should be returned for service. There are 
no user alignments possible. 
(Written Calibration Procedures are 
available from DeFelsko at no charge, 
and are included on the PosiSoft CD with 
every Memory model). 
Calibration Standards:
To calibrate the 6000 you must have a set 
of Thickness Calibration Standards with 3 
to 5 values that cover the full range of the 
probe being used. These standards must 
have been measured to an accuracy that 
is four or more times better than the 
accuracy of the Gage at the thickness 
being measured. 
Example: If the gage has an accuracy of
±
2 microns at the thickness being
measured, then the standards have to 
have been verified to an accuracy of 
better than or equal to 
±
0.5 micron.
(Traceable Calibration Standards are 
available for each 6000 model. Contact 
your dealer for details). 
