Inch/Metric Conversion

(Config Menu)

Selecting the

MILS?

MILS? or MICRONS?

MICRONS?

Config Menu option converts the display
and all stored readings from inch to metric
or vice versa.

RESET Feature

It may be desirable to return the 6000 to a
“known” state when settings have been
changed purposely or inadvertently, if the
gage behaves in an unfamiliar or
unexpected manner, or if an object of
known thickness is not available to adjust
gage readings.

There are two types,

RESET

RESET and

XRESET

XRESET.

NOTE:

Keep the probe away from metal

during

RESET

RESET or XRESET

XRESET.

RESET

(Main Menu)

Selecting

RESET

RESET returns the gage to a

basic condition without erasing any
readings stored in gage memory. The
following changes are made to the gage
after selecting

RESET

RESET:

all MODES are turned off
(memory, statistics, HiLo).

a zero reading and a

symbol

appear on the display. The

symbol

disappears if the Gage is zeroed or
optimized.

RESET

RESET is handy when:

you wish to return the gage to a basic
condition, but you do not wish to lose
stored readings and Config Menu
settings