Hilo mode – DeFelsko PosiTector 6000 v.4.0 User Manual
Page 18
18
•
another reading is taken, in which case
the sequence starts over.
•
the 8
8 button is pressed, in which case
the last reading is displayed.
H
I
L
O
Mode must be off to activate
S
TATISTICS
Mode. Adjustments cannot be
made to readings while in S
TATISTICS
Mode. To turn S
TATISTICS
Mode off,
select
STATOFF?
STATOFF?
HiLo Mode
(Main Menu)
Standard (2) and Memory (3) models only
When
HiLo ON?
HiLo ON? is selected, the
current
Lo
Lo setting is displayed. Adjust
down 8
8 or up 9
9. Alternatively, measure a
coating with a thickness close to the
required value and make final
adjustments with the 8
8 9
9 buttons. Press 8
8
9
9 simultaneously to accept this value.
The last
Hi
Hi setting is now displayed.
Follow the same procedure to adjust the
Hi
Hi setting. The
symbol will indicate
HiLo
HiLo Mode is now ON.
Measurements will now be compared to
your defined limits. The Gage BEEPS
TWICE (normal) if results are within those
limits and ONCE if outside. The single
BEEP is a LOW tone with a
æ if it is
below the Lo limit, and a HIGH tone with
a
ä if it is above the
Hi
Hi limit.
NOTE:
The LED will only BLINK if
readings are within limits.
S
TATISTICS
Mode must be off. Calibration
adjustments cannot be made while in
H
H
II
L
L
O
O
M
ODE
. To turn
H
H
II
L
L
O
O
M
ODE
off
select "
HiLoOFF?
HiLoOFF?".