Changing The Battery
As the batteries become weak the
symbol will appear during
measuring. The Gage can still be used in
this condition but the batteries should be
changed at the earliest opportunity.
USE ONLY AA ALKALINE BATTERIES
Nickel-cadmium and nickel-metal hydride
rechargeable batteries will work but the
low battery symbol may stay on.
To retain all user settings and stored
memory readings, disconnect the battery
only after the Gage has automatically
powered-down.
Battery Life: 50 hours continuous or
36,000 readings
Trouble Shooting
If “
SERVICE
SERVICE” appears on the display,
the Gage must be returned to the
manufacturer or dealer for repair.
A “
PWR LOST
PWR LOST” condition may appear if
the batteries are disconnected while the
Gage is on. It may also indicate a low
battery condition. Zero adjustments will
have to be re-done and memory may be
erased. If “
PWR LOST
PWR LOST” persists, change
the batteries and perform an
XRESET
XRESET.
As with any electronic device, harsh
temperature and electromagnetic
environments or rough handling may alter
the performance of some Gage
components causing the Gage to exhibit
unusual behavior. An
XRESET
XRESET will
enable the Gage to recover from many of
these disturbances without the necessity
of returning the Gage for service.