Changing The Battery

As the batteries become weak the

symbol will appear during

measuring. The Gage can still be used in
this condition but the batteries should be
changed at the earliest opportunity.

USE ONLY AA ALKALINE BATTERIES

Nickel-cadmium and nickel-metal hydride
rechargeable batteries will work but the
low battery symbol may stay on.

To retain all user settings and stored
memory readings, disconnect the battery
only after the Gage has automatically
powered-down.

Battery Life: 50 hours continuous or

36,000 readings

Trouble Shooting

If

SERVICE

SERVICE appears on the display,

the Gage must be returned to the
manufacturer or dealer for repair.

A

PWR LOST

PWR LOST condition may appear if

the batteries are disconnected while the
Gage is on. It may also indicate a low
battery condition. Zero adjustments will
have to be re-done and memory may be
erased. If

PWR LOST

PWR LOST persists, change

the batteries and perform an

XRESET

XRESET.

As with any electronic device, harsh
temperature and electromagnetic
environments or rough handling may alter
the performance of some Gage
components causing the Gage to exhibit
unusual behavior. An

XRESET

XRESET will

enable the Gage to recover from many of
these disturbances without the necessity
of returning the Gage for service.