beautypg.com

DeFelsko PosiTector 6000 v.4.0 User Manual

Page 19

background image

19

Memory Mode

(Main Menu)
Memory (3) models only

When

MEM

MEM

_

ON?

ON? is selected the

symbol and a counter (

0

0) appear. 5000

readings in up to 100 batches (groups)
can be recorded.

In M

EMORY

M

ODE

readings are

simultaneously displayed and stored. The
current BATCH number (

B1

B1) is displayed.

Remove the last reading from memory by
pressing 8

8. Erase ALL stored readings by

selecting "

CLR

CLR MEM?

MEM?". Calibration

adjustments cannot be made while in
M

EMORY

Mode. To turn M

EMORY

Mode off

select "

MEM OFF?

MEM OFF?”.

NOTE::

Press 9 at any time for a

summary view of the average,
standard deviation, number of
readings, minimum thickness reading
and maximum thickness reading in the
current batch. Then press 8 or take a
reading to exit the summary view.

Batch Markers

Readings taken for different jobs or parts
can be grouped together or "separated"
by placing Batch Markers between sets of
measurements. Select

B MARK?

B MARK? from

the Main Menu after a series of readings.
A Batch Marker is set automatically if you
turn H

I

L

O

or S

TATISTICS

Modes on or off.

Statistical summaries are calculated for
each batch during downloading.

Memory AND Statistics Modes

Together

When both these MODES are ON,
individual measurements ARE NOT
STORED. Instead, when 9

9 is pressed, the