DeFelsko PosiTector 6000 v.4.0 User Manual
Page 4

4
Power-up / Power-down
The 6000 powers-up when any button is
depressed and displays the last
measurement. Built-In probe gages turn
on automatically when the probe touches
the surface.
To preserve battery life, the gage powers-
down after 60 seconds of no activity.
There is no OFF button. However, to
quickly power-down, press both buttons
simultaneously and hold until gage
displays “
OFF
OFF”.
All settings are retained during power-
down including the last measurement.
Always be sure the gage is powered
down before changing the batteries.
How To Take A Measurement
1. On Built-In probe models, slide the
probe cover open FULLY. It acts as a
second "foot" to steady the Gage.
2. Turn gage on with any button. Built-In
probe models turn on automatically
when the probe is placed on a surface.
3. Place the probe FLAT on the surface to
be measured and HOLD STEADY.
When a valid measurement is
calculated, the Gage BEEPS, BLINKS
twice and displays the measurement.
4. Lift probe AT LEAST 2 INCHES (50
mm) from the surface between
measurements - OR - leave probe in
the same location for continuous
measurements about every 5 seconds.
5. When finished, slide the Built-In probe
cover FULLY shut to prevent Gage
from accidentally turning on and
draining the battery. Keep Separate
probes away from metal so they can
power down automatically.