DeFelsko PosiTector 6000 v.4.0 User Manual
Page 10

10
Calibration Interval
User specific intervals should be 
established based upon frequency of use, 
handling practices and measuring 
environment. 
Effects of Temperature
If measurements will be made in extreme 
hot or cold temperatures, measure your 
Calibration Standards in the working 
environment. The 6000 should then be 
optimized in that same environment. 
How to Optimize
One- or Two-point Optimization should be 
performed if readings are not falling within 
the expected range of thickness for the 
application being measured. 
One-point Optimization
Simple Zero Adjustment
(F, FS, FRS, N, NS, NRS, FN, FNS, FNRS, SPFS and 
NAS probes only) 
Measure your uncoated part. If the Gage 
does not read “
0
0” within the tolerance of
the probe being used, lift the probe from 
the surface and adjust the display down 8
8
or up 9
9 until the display reads “0
0”.
Measure and adjust until the average of a 
series of readings on the uncoated 
surface is “
0
0” (within tolerance).
Average Zero Adjustment
(Main Menu)
On rough or curved surfaces the Simple 
Zero Adjustment method must be 
repeated several times to get a good “
0
0”.
A preferred method is to take several 
readings on the uncoated part and 
average the result. 