DeFelsko PosiTector 6000 v.4.0 User Manual

Page 14

background image

14

XRESET

(Config Menu)

Selecting

XRESET

XRESET returns the gage to

an out-of-the-box condition by performing
a RESET and restoring factory settings.
The

XRESET

XRESET function makes all of the

same changes as

RESET

RESET, and also

make the following changes:

all readings in memory are erased.

measuring units, power down, button
lock and resolution settings are
restored.

XRESET

XRESET is handy when:

attempting to correct unexpected gage

behavior

you are uncertain what previous
adjustments have been made and wish
to return the Gage to a known
condition.

FN Combination Probes

FN probes combine the capabilities of
both "F" and "N" probes.

The probe first attempts a measurement
using the magnetic principle. If the
coating is non-magnetic, within the
probe’s thickness range and the substrate
is ferrous, a reading is displayed with the
letter "F".

If no valid reading is obtained with the
magnetic principle the probe
automatically attempts a measurement
using the eddy current principle. If the
coating is non-conductive, within the
probe’s thickness range and the substrate
is a non-ferrous metal, a reading is
displayed with the letter "N".

If a reading is still not obtained, dashes
( -- -- -- -- -- ) appear on the display.