Two-point Optimization

(FT, FTS, NT, NTS, FNT, FNTS, FHS, NHS and
Microprobes only)

This method requires taking a reading on
both the uncoated object (

ZERO

ZERO) and on

a known thickness (

2ND PNT

2ND PNT). This

second point should be near the top end
of your application thickness to be
measured.

1.

ZERO?

ZERO? the Gage as described in

Average Zero Adjustment” (p.10).

2. Select the

2ND PNT?

2ND PNT? (Main Menu)

option.

3. Press 9

9 to set the number of readings

used to obtain an average, typically 3
to 9. The greater the variation between
readings, the more readings should be
taken to obtain an average. The
display will show the number of
readings remaining to be taken; i.e.

2NDPNT4X

2NDPNT4X

4. Repeatedly measure the known

thickness (coated part, a plastic shim
or a Coating Thickness Standard) until
the display shows

2PT

2PT=value. The

Gage will display “

WAIT

WAIT” for 2 seconds

between readings to allow the user to
correctly position the probe on the
surface.

5. Lift the probe from the surface and

adjust the displayed reading down 8

8 or

up 9

9 to the known thickness value.

Press 8

8 9

9 simultaneously to accept this

value.