5 seca 360°wireless printer – Seca 515 User Manual
Page 70

70 •
13.5 seca 360°wireless printer
Problem
Cause
Remedy
No printout
The
seca 360°
wireless printer is not
switched on.
Switch on
seca 360°
wireless printer
No seca wireless network has been
established between the device and
the
seca 360°
wireless printer.
Establish seca wireless network
No patient data on the printout
On repeat measurement: No patient
assigned
Call up the
patient
tab and assign patient
(see “Searching for patient data” on
page 37)
On initial measurement: Patient not yet
created in database
Call up the
patient
tab and create the
patient (see “Creating new patients” on
page 39)
No BIA results on the printout
No BIA measurement was performed
Call up the
bia
tab and perform the
measurement (see “Performing bioelectric
impedance analysis (BIA)” on page 31)
No body size on the printout
No body size has been entered.
Call up the
weight/height
tab and enter the
body size (see “Manually entering body
size” on page 29)
No body size has been sent from the
seca 360°
length measuring device.
Measure the body size of the patient
again, then press the
send/print
button on
the length measurement device (see
“Transmitting body size via seca wireless
network” on page 29)