2 establishing power supply, 3 setting up device, 2 establishing power supply 6.3 setting up device – Seca 515 User Manual
Page 19

Before you really get started... • 19
English
6.2
Establishing power supply
WARNING
!
Using the wrong power supply may cause bodily injury or
damage to the equipment
Conventional power supplies may provide a higher voltage than
is indicated on them. The device may overheat, catch fire, melt,
or short circuit.
− Use only the original seca plug-in power supply as
contained in the scope of delivery and listed in section
”Spare parts” on page 73.
The connection for the power pack is located on the underside of the
weighing platform. To establish the power supply, proceed as follows:
1. Insert the the mains plug required for your power supply into the
power pack.
2. Tilt the device forward.
3. Insert the device plug of the power pack into the connector socket of
the device.
4. Carefully return the device to an upright position.
5. Plug the power pack into a mains socket.
6.3
Setting up device
The device is fully assembled upon delivery.
ATTENTION!
Incorrect measurement due to force shunts
If the device with the housing is placed on top of something,
e.g. due to an uneven or soft floor covering, the weight will not
be correctly measured.
− Set the device up so that only its foot screws are in
contact with the ground.
1. Place the device on firm and even ground.
ATTENTION!
Incorrect set-up may cause incorrect measurements
The spirit level is very sensitive. Additional weights, such as
hand towels, can result in incorrect scale calibration.
− Make sure the scale is calibrated without any additional
weight.
2. Calibrate the device by turning the foot screws.
The air bubble in the spirit level must be located in the center of the
circle.
Spirit level
Air bubble