Measurement procedure

The operating concept is based on the following typical measurement
procedure:

• Measure weight and height

• Perform bioelectric impedance analysis

• Assign measurements to a patient

• Evaluate measured results

• Save measurement procedure

The order of the tabs on the touchscreen display follows this sequence. It
is possible to operate in a different order.

ATTENTION!
Data loss
If there is no new entry for approx. 5 minutes, the current
measurement is discarded. In this case, the complete
measurement procedure must be repeated.
− Save or print the measured results immediately after the

conclusion of the measurement procedure (see “Closing the
measurement procedure” on page 42).

Automatic standby

The device automatically switches to standby if there are no entries on the
device for 5 minutes. This has the following effects:

• Measured results and settings that have not been saved are lost.

• The LED of the ON/OFF button turns green.

• The touchscreen display goes out.

Switching off the device

WARNING!
Electric shock
The device cannot be put out of operation by pressing the ON/OFF
button.
− Always pull out the mains plug if the device is to be put out of

operation, e.g. for hygienic preparation.

♦ Give the ON/OFF button a brief press.

The LED of the ON/OFF button turns green. The touchscreen display
goes out. The device is in standby.

♦ Give the ON/OFF button a long press

The LED of the ON/OFF button goes out. The touchscreen display
goes out. The device is switched off.

NOTE:
When switching on from standby, the device starts in the

weight/

height

tab. When switching on a device that is switched off, the

internal PC boots up again. This takes several seconds.