XJ
Selection of reference junction
compensation
Mode
Setup mode
Setting
XJp1, p2, p3
p1
Channel number (001 to 560)
p2
Selection of reference junction
compensation value
INT
Internal compensation circuit
EXT External junction compensation
p3
External reference junction compensation
value
(-20000 to 20000)
Example
Set channel 01 of subunit 0 to external junction
compensation at a compensation value of 1000
µ
V.
XJ001, EXT, 1000
Comments
• For p2 = INT, parameter p3 is ineffective.
• Unit of p3 is
µ
V.
XG
Sets computation error handling.
Mode
Operation mode
Setting
XGp1, p2, p3, p4, p5
p1 Computation error handling (+OVER/-
OVER)
p2 Scale unit for TLOG computation (OFF, /
SEC, /MIN, /HOUR)
p3 Handling of abnormal data in a channel for
TLOG computation
ERROR: Handled as computation error.
SKIP:
Abnormal data are skipped
(ignored) and computations are
executed.
p4 Handling of overflow data in a channel for
TLOG computation
ERROR: Handled as computation error.
SKIP:
Abnormal data are skipped
(ignored) and computations are
executed.
LIMIT: If linear scaling has been set, its
upper- and lower-limit values
are computed. If it has not been
set, the upper- and lower-limit
values in the measurement
range are computed.
p5
handling of data for TLOG.PSUM(only
for PULSE input module)
OVER A result of the computational
expression TLOG.PSUM
(XXX) exceeding 99999999 as
an overflow
ROTATE
A result of the computational
expression TLOG.PSUM
(XXXX) exceeding 99999999
to continue computing with the
value following 99999999
reset to 0.
.
Example
Compute computation error as +OVER and TLOG
computation scale value as off, and ignore
abnormal data in a channel and overflow data in a
channel for computation and a result of the
computational expression TLOG.PSUM (XXX)
exceeding 99999999 as an overflow.
XG+OVER, OFF, SKIP, SKIP, OVER
Comments
• This command only works with the optional
computation function or if a pulse module is
installed.
• p2 is effective for the totalization of flow
signals, which are expressed in engineering
units /s, /min, /h.
If p2 is set according to the input unit, the
measurement data are computed based on that
unit at the specified measurement intervals. For
example, set the measurement interval to 2s,
the input value to 100 m3/min, and p2 to /MIN.
By doing this, because 2s/60s is multiplied for
each measurement interval, then after 1 minute,
approximate actual input values are obtained.
VS
Selects the relay that can be turned ON/
OFF externally.
Mode
Setup mode
Setting
VSp1, p2
p1 Unit number
I
DA100 main unit
S
Internal switch
0 to 5
Sub units 0 to 5
p2 Number of relay that can be turned ON/Off
externally
NONE
No relay
01 to 10
Relay numbers 01 to 10
01 to 20
Relay numbers 01 to 20
01 to 30
Relay numbers 01 to 30
01 to 40
Relay numbers 01 to 40
01 to 50
Relay numbers 01 to 50
01 to 60
Relay number 01 to 60
Example
Specify the relays from the first of slot 0 to the
10th of slot 2 in sub unit 0 to those which can be
turned ON/OFF externally.
VS0, 01-30
Comments
• p2 is set with number of the first relay (fixed to
01), and that of the last relay.
• The relays specified here can only be turned
ON/OFF by using the VD command.
• Configurations such as alarm reflash, AND/OR,
and hold are ineffective for the relays specified
here.
XE
Establishes content of the setup mode
setting.
Mode
Setup mode
Setting
XEp1
p1
Selection of establishment or destruction
STORE
Establishment
ABORT
Destruction
Example
Store a parameter set in the setup mode in
NVRAM.
XESTORE
Comments
• A parameter set in the setup mode becomes
ineffective in its setting if the mode is changed
without STORE. After setting all parameters
in the setup mode, be sure to store the set data
in the internal memory using the XE command.
In addition, after normal processing with the
XE command, the mode is transferred to the
operation mode.
• Since execution of the XE command takes an
indefinite time, return an ACK after the
processing is completed. On the controller side,
execute the next processing after receiving an
ACK after transmitting the XE command. The
ACK status format is as follows:
6.1 Setting Command