< 4. OPERATION >
4-6
IM 12T03A01-02E
Function No. ‘7.’: Sets high limit of the output range
Sets a high limit density value (g/cm
3
) of an analog output range.
(Note) Difference between high and low limit value should be 0.05 g/cm3 or more. If an
inadequate value is set, the function No. returns to ‘6.’ operations of ‘6.’ and ‘7.’ are related;
when setting is made at ‘6.’ setting should also be made at ‘7.’.
Function No. ‘8.’: Sets reference temperature of measuring liquid
Sets a reference temperature (°C) of measuring solution.
Function No. ‘9.’: Sets temperature coeffi cient of measuring liquid
Sets the value which is -1 x 10
5
(= -100000) times the temperature coeffi cient of measuring
solution (g/cm
3
/°C) to change it to plus integer.
(Note) Purifi ed water is an exception; its maximum density is at approx. 4 °C.
Generally, temperature coeffi cients of any liquids are minus number.
Function No. ‘A.’: Sets detector constant A
Sets the detector constant A. Since the detector constant A is an inherent characteristic of the
detector, this is indicated in each detector.
(Note) Detector constant A changes with each calibration.
Function No. ‘B.’: Sets detector constant B
Sets the detector constant B. Since the detector constant B is also an inherent characteristic of
the detector, this is indicated in each detector.
Function No. ‘C.’: Displays detector calibration coeffi cient C (span)
Indicates either the one-point or two-point calibration has been performed. 1 is indicated for one-
point calibration and calibration coeffi cient (span) for twopoint.
Function No. ‘D.’: Displays detector calibration coeffi cient D (zero)
Indicates either the one-point or two-point calibration has been performed. 0 is displayed for
one-point calibration, and calibration coeffi cient (zero point) than 0 is displayed for two-point
calibration.