< 1. GENERAL >
1-2
IM 12T03A01-02E
1.1.2 Detector
(1) General Purpose Detector Model VD6D
Detector construction: Non-explosion protection, rain-proof construction
Case material:
Cast Aluminum alloy
Case coating:
Epoxy resin, baked fi nish
Case color:
Jade green (equivalent to Munsell 7.5BG4/1.5)
Wetted part materials:
Base:
SUS316
Vibrator:
SUS316 or Ni (Au Brazing: BAu·4)
Measuring liquid temperature:
–10 to 100 °C
Measuring liquid pressure:
2 MPa G or less
Maximum pressure:
4.9 MPa G
Steam tracing:
Available
Process connection:
Rc1/4
Electrical connection:
G3/4
Mounting:
2-inch. pipe mounting
Ambient temperature: –10 to 50 °C
Weight:
Approx. 12 kg
(2) Flameproof (Explosionproof) Detector Model VD6DF
Detector construction: TIIS; d2G3 or FM; Class I, Division 1, Groups C and D, Flameproof
construction
Process connection:
Rc1/4 or 1/4NPT female (only for VD6DF-*B/FM)
Electrical connection:
G3/4 or 3/4NPSM female (only for VD6DF-*B/FM)
Specifi cations are the same as for the (1) General Purpose Detector except for the above
construction.
(3) Sanitary Use Detector Model VD6DS
Process connection:
Special joint for connection to JIS 6A (6 mm dia.) pipe (within gasket)
Wetted part materials: Added to the standard model
Gasket:
Tefl on
O-Ring:
Viton
Stream tracing:
Not available
Specifi cations are the same as for the (1) General Purpose Detector except for the above two items.
Temperature detector protecting tubes are detachable.
Note: These detectors cannot be used with highly corrosive liquids and solutions likely to stick
to sensors. If it is desired to be applied to solutions containing slurry or sludge, consult with
YOKOGAWA. For measuring NaOH solutions, use sensors with a nickel vibrator.