3 converter model dm8c, Converter model dm8c -3 – Yokogawa DM8C/VD6 Liquid Density Analyzer User Manual
Page 13

< 1. GENERAL >
1-3
IM 12T03A01-02E
1.1.3 Converter Model DM8C
Display: Digital
display,
fi ve digits LED
Display contents:
Density
(g/cm
3
) after conversion to reference temperature (center temperature)
Density
(g/cm
3
) at the measuring temperature
Measuring liquid temperature (°C)
Set density value for the calibration liquid (g/cm
3
) (displayed on call)
Temperature
coeffi cient set value for the calibration liquid (x10
–5
g/cm
3
/°C) (displayed on call)
Output signal set value (%) (displayed on call)
Setting for output range low limit (g/cm
3
) (displayed on call)
Setting for output range high limit (g/cm
3
) (displayed on call)
Reference temperature (center temperature) set value (°C) (displayed on call)
Temperature
coeffi cient set value for the measuring liquid (x10
–5
g/cm
3
/°C) (displayed on call)
Fault contents display
Output signal:
Analog output: 4 to 20 mA DC (load resistance 550
Ω or less), and 0 to 1 V DC (load
resistance 250 k
Ω or more), isolated output.
Density
(g/cm
3
) after conversion to the reference temperature
Digital output:
RS-232-C
(1) Communication
specifi cations
• Asynchronous
system
Start bit; 1 bit, Stop bit; 2 bit, Parity; none
• Baud rate; 1200 bps
• Data format; ASCII, Data length; 8 bit
• Wiring system; Two-wire system (output only)
(2) Output
signal
• Transmitting
contents
Measured data; Density (g/cm
3
) after conversion to the reference temperature, density (g/cm
3
) at
the measured temperature, measured liquid temperature (°C)
Calibration state; calibration start, error No., calibration end
Failure alarm; error No.
• Signal
level
Output voltage; ON; 9 ±3 V, OFF; -9 ±3 V
Output impedance; 300
Ω