Yokogawa DM8C/VD6 Liquid Density Analyzer User Manual
Page 14

< 1. GENERAL >
1-4
IM 12T03A01-02E
• Output
format
Measured
data
space
space
# * . * * * * _ * . * * * * _ * * * . *
C
R
L
F
(Note) Data are output in the order of density (data converted into that at a reference
temperature), density (at measuring temperature), and measuring liquid temperature.
Hold data (data immediately before entering the mode) is output during the time of
calibration or maintenance mode. However, holding is released immediately after
calibration or maintenance mode.
Calibration
status
space
# CALIBRATION _ START
C
R
L
F
(Note) Nothing is output during calibration. However, when a parameter error occurs, output
becomes as follows.
# ERROR - *
C
R
L
F
(*: 5 or 6)
space
# CALIBRATION _ END
C
R
L
F
Fail
alarm
space
# ERROR - * _ ERROR - *
C
R
L
F
(Note) When multiple errors occur, each error No. for the respective errors is output.
space
Example: # ERROR - 1 _ ERROR - 3
C
R
L
F
Output signal span:
0.05 to 0.5 g/cm
3
settable
Reference temperature set range:
0 to 100 °C (in increments or decrements of 1 °C)
Contact output on failure: One point. Contact closed on failure or power failure. Contact open
when
normal.
Permissible voltage:
220 V DC, 250 V AC
Permissible current:
2A (resistive load)
Permissible contact power:
60 W
Fault detecting contents: Detector failure and converter failure
Failure output:
Analog signal: Falls down to about –10 % of the output signal span
Digital signal:
Error message outputs
Output signal hold:
Holds in the CAL. or Maintenance mode.
Settable range for temperature coeffi cient:
0 to 0.002 g/cm
3
/°C
Calibration procedure: One-touch calibration by strong calibration liquid density
(onepoint
calibration)
Ambient temperature: –10 to 55 °C
Power supply:
90 to 132 V AC or 180 to 264 V AC, 50/60 Hz