Output stop – Yokogawa DM8C/VD6 Liquid Density Analyzer User Manual
Page 46

< 4. OPERATION >
4-4
IM 12T03A01-02E
Digital signal (Note 1)
Measuring mode (DSPL)
Measuring mode (DSPL)
Maintenance mode (1)
(MAINT)
Maintenance mode (3)
(MAINT)
Transfer to
Calibration mode
(1)
(2)
(3)
While in calibration
computation
Calibration mode
function No. transfer
or waiting for mode
transfer
Calibration mode (CAL)
Function No.
transfer (2)
Calibration start
Calibration end
Output stop
Measured
value
Hold
Hold
(Note2)
Measured
value
Measured
value
Measured
value
Measured
value
Hold
Analog signal
(1)
(2)
(3)
Measured
value
Hold
Hold
Measured
value
Measured
value
Measured
value
Hold
Note 1:
The density value after temperature conversion, density value at the measuring temperature and measuring liquid
temperature are transmitted in digital form. (Only the density value after temperature conversion is transmitted in analog form.)
Note 2:
The density converter displays CAL-1 or CAL-2.
Figure 4.3
Digital and Analog Signal Output Statuses
Eight function Nos. from ‘3’ to ‘A’ are provided in the calibration mode (see Table 4.1). Among
them, ‘6’ and ‘A’ are to specify the calibration computation and others are to input data required
for the calibration.
Function No. ‘3’: Sets density of calibration solution 1
Density value (g/cm
3
) of a calibration solution for one-point calibration or 1st calibration of two-
point calibration can be input. The input density is a data at the reference temperature.
(Note) If relation between temperature and density is well known, a density at a temperature
can be set. However, the density data at the reference temperature is recommended to
achieve accurate calibration.
Function No. ‘4’: Sets temperature of calibration solution 1
Input the temperature of the calibration solution 1 of which density at the reference temperature
is set at ‘3’.
Function No. ‘5’: Sets temperature coeffi cient of calibratio solution 1
Enter the value which is -100000 times the temperature coeffi cient (g/cm
3
/°C) of the liquid for
calibration 1.
(Note) The temperature coeffi cient -0.00086 g/cm3/°C becomes the value 086.
Function No. ‘6’: Starts calibration 1 (press SET key)
Pressing “SET” key executes calibration computation, or also executes the fi rst calibration of the
two-point calibration.
(Note) For calibration procedure, refer to Section 4.4.