Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual
Page 76

4-18 Profibus
IM 12A00A01-61E
4-4-3-4. Function Block Parameters dO202 (continued)
Parameter
default
Alternatives
Unit r/W Slot, Index data Type
(byte,bit) (bytes)
Manual temperature
logbook1 (2)
not logged (1), logbook2 (3)
R/W 4,78
unsigned8 (1)
Manual calibration (0%)
logbook1 (2)
not logged (1), logbook2 (3)
R/W 4,83
unsigned8 (1)
Manual calibration (100%) logbook1 (2)
not logged (1), logbook2 (3)
R/W 4,79
unsigned8 (1)
Air calibration (0%)
logbook1 (2)
not logged (1), logbook2 (3)
R/W 4,92
unsigned8 (1)
Air calibration (100%)
logbook1 (2)
not logged (1), logbook2 (3)
R/W 4,80
unsigned8 (1)
H
2
O calibration (0%)
logbook1 (2)
not logged (1), logbook2 (3)
R/W 4,82
unsigned8 (1)
H
2
O calibration (100%)
logbook1 (2)
not logged (1), logbook2 (3)
R/W 4,81
unsigned8 (1)
Zero setting changed
logbook1 (2)
not logged (1), logbook2 (3)
R/W 4,84
unsigned8 (1)
Sensitivity changed
logbook1 (2)
not logged (1), logbook2 (3)
R/W 4,90
unsigned8 (1)
Delta t for calibation
changed
not logged (1)
logbook1 (2), logbook2 (3)
R/W 4,86
unsigned8 (1)
Delta PV for calibration
changed (0%)
not logged (1)
logbook1 (2), logbook2 (3)
R/W 4,85
unsigned8 (1)
Delta PV for calibration
changed (100%)
not logged (1)
logbook1 (2), logbook2 (3)
R/W 4,91
unsigned8 (1)
Salinity compensation
changed
not logged (1)
logbook1 (2), logbook2 (3)
R/W 4,87
unsigned8 (1)
%Saturation pressure
changed
not logged (1)
logbook1 (2), logbook2 (3)
R/W 4,88
unsigned8 (1)
Calibration pressure
changed
not logged (1)
logbook1 (2), logbook2 (3)
R/W 4,89
unsigned8 (1)
New sensor installed
logbook1 (2)
not logged (1), logbook2 (3)
R/W 4,93
unsigned8 (1)