Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual
Page 31

IM 12A00A01-61E
3-26 Foundation Fieldbus
DEVICE_STATUS_2 PH202
Hexadecimal Display through DD
0x80000000
0x40000000
0x20000000
0x10000000
0x08000000
0x04000000
0x02000000
0x01000000
0x00800000
0x00400000
0x00200000
0x00100000
0x00080000 FF interface checksum error
0x00040000 EXA checksum error (E21)
0x00020000 Internal communication failure
0x00010000 FF interface eeprom failure
0x00008000 EXA eeprom failure (E20)
0x00004000 mismatch between FF- and EXA parameter
0x00002000
0x00001000
0x00000800
0x00000400
0x00000200
0x00000100 calibration timer expired (E16)
0x00000080 reference impedance exceeds low limit (E4.2)
0x00000040 reference impedance exceeds high limit (E5.2)
0x00000020 glass impedance exceeds low limit (E4.1)
0x00000010 glass impedance exceeds high limit (E5.1)
0x00000008 temperature sensor shorted (E8)
0x00000004 temperature sensor open (E7)
0x00000002 tertiary value exceeds limits (E12)
0x00000001 primary value exceeds limits (E9)
DEVICE_STATUS_2 SC202
Hexadecimal Display through DD
0x80000000
0x40000000
0x20000000
0x10000000
0x08000000
0x04000000
0x02000000
0x01000000
0x00800000
0x00400000
0x00200000
0x00100000
0x00080000 FF interface checksum error
0x00040000 EXA checksum error (E21)
0x00020000 Hart communication failure
0x00010000 FF interface eeprom failure
0x00008000 EXA eeprom failure (E20)
0x00004000 mismatch between FF- and EXA parameter
0x00002000
0x00001000
0x00000800
0x00000400
0x00000200
0x00000100 matrix error (E4)
0x00000080 concentration table error (E18)
0x00000040 conductivity exceeds usp limit (E13)
0x00000020 polarization detected (E1)
0x00000010 temperature compensation error (E2)
0x00000008 temperature sensor shorted (E8)
0x00000004 temperature sensor open (E7)
0x00000002 conductivity exceeds low limit (E6)
0x00000001 conductivity exceeds high limit (E5)