Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual
Page 44

IM 12A00A01-61E
Foundation Fieldbus 3-39
3-39
Index Parameter name
Factory
Valid
Description
Default
Range
2035 COMP_METHOD
NaCl
NaCl, TC, matrix
Method of process temperature compensation for the primary value
2036 COMP_MATRIX_SEL
H
2
SO
4
H2SO4, 0 -100ºC, 0 - 5%, When matrix compensation is required one can
H2SO4, 0 -100ºC, 2.5 - 25%, make a selection out of 8 predefined matrices and
HCl, 0 - 60ºC, 0.5 - 5%,
one user definable matrix
HNO3, 0 - 80ºC, 0.5 - 5%
HNO3, 0 -80ºC, 2.5 - 25%
NaOH, 0 -100ºC, 0.5 - 5%
NaOH, 0 -100ºC, 0.5 - 15%
User programmable matrix
2037 TERTIARY_COMP_METHOD NaCl
NaCl, TC, matrix
Method of process temperature compensation
for the second conductivity value
2038 TERT_TEMPERATURE
2.1
0 ~ 3.5 %/ºC
Process temperature compensation factor for the
_COEFF
second conductivity value
2039 ALARM_SUM
-
2040 DEV_ALARM
-
Device Alarm is used to give the status of the analyser.
See separate table for error messages
2041 LOGBOOK1_RESET
Idle
Idle, Reset
Reset the pointer to the first (oldest) event in logbook 1
2042 LOGBOOK1_EVENT
-
Event whereto the pointer is referenced. When parameter
is read, the pointer is increased by one.
2043 LOGBOOK2_RESET
Idle
Idle, Reset
Reset the pointer to the first (oldest) event in logbook 2
2044 LOGBOOK2_EVENT
-
Event whereto the pointer is referenced. When parameter
is read, the pointer is increased by one
2045 LOGBOOK_CONFIG
-
Per event one can decide whether it should be logged and
in which logbook (1 or 2) it should be logged
2046 TEST_1
-
....
-
2047 to 2057 are, like 2046 and 2058, service parameters
2058 TEST_13
-
2059 CALIB_SENSOR_CONST
1.88
0.2~19.99 cm
-1
The calibrated cell constant [1/cm].
2060 MATRIX_TEMP_RANGE
-20~140ºC, 0~280ºF
The temperature values of the user defined matrix.
2061 SOLUTION_1
0~1.999 S/cm
The conductivity values of solution 1.
2062 SOLUTION_2
0~1.999 S/cm
The conductivity values of solution 2.
2063 SOLUTION_3
0~1.999 S/cm
The conductivity values of solution 3.
2064 SOLUTION_4
0~1.999 S/cm
The conductivity values of solution 4.
2065 SOLUTION_5
0~1.999 S/cm
The conductivity values of solution 5.
2066 CONCENTRATION_
disabled
0=Not initialized, 1=disabled,
Enabling/Disabling of concentration measurement.
MEASUREMENT
2=enabled
2067 CONCENTRATION_0
0
0~100
The concentration [%] corresponding to the 0 percent table value
2068 CONCENTRATION_100
100
0~100
The concentration [%] corresponding to the 100 percent table value
2069 CONC._TABLE_LOW
0~1.999 S/cm
The first part of the concentration table (0 till 30 percent).
2070 CONC._TABLE_MID
0~1.999 S/cm
The second part of the concentration table (35 till 65 percent).
2071 CONC._TABLE_HIGH
0~1.999 S/cm
The third part of the concentration table (70 till 100 percent).
2072 E5_LIMIT
3 (3 S)
0~5 S
The upper range accuracy limit of the analyzer [Siemens].
2073 E6_LIMIT
5e-6 (5
μS) 0~5 S
The lower range accuracy limit of the analyzer [Siemens].
2074 DISPLAY_RESOLUTION
1
0 to 8
Display resolution, 0, Not initialized (can not change)
1, Auto Ranging, 2, x.xxx μS/cm, 3, xx.xx μS/cm,
4, xxx.x μS/cm, 5, x.xxx mS/cm, 6, xx.xx mS/cm,
7, xxx.x mS/cm, 8, xxxx mS/cm
2075 PASSCODE_MAINT
0
0, 111, 333, 777,
Passcode used to protect the maintenance menu
888, 123, 957, 331, 546, 847
2076 PASSCODE_COMM
0
0, 111, 333, 777,
Passcode used to protect the commissioning menu.
888, 123, 957, 331, 546, 847
2077 PASSCODE_SERVICE
0
0, 111, 333, 777,
Passcode used to protect the service menu.
888, 123, 957, 331, 546, 847
2078 ERROR_CONFIG*
0=soft, 1=hard
Soft/Hard fail configuration (checked = hard unchecked = soft).
2079 CONFIGURATION*
autoreturn on
Instrument specific configuration. See below for correct
bitnr. setting
2080 TRANSMITTER_TIME
Transmitter time.”
2081 TEST_14
Service parameter.