6-3-2 transducer block sc202, 6-3-2 transducer block sc202 -35 – Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual
Page 40

IM 12A00A01-61E
Foundation Fieldbus 3-35
3-35
3-6�-3-2 Transducer Block SC202
Index Parameter name
Factory
Valid
Description
Default
Range
2000 BLOCK HEADER
TAG: “TB”
General information about thefunction block
2001 ST_REV
-
-
The revision level of the static data associated with the function
block. The revision value will be incremented each time a static
parameter value in the block is changed
2002 TAG_DESC
“ “
The user description of the intended application of the block
2003 STRATEGY
1
The strategy field can be used to identify grouping of blocks.
This data is not checked or processed by the block
2004 ALERT_KEY
1
The identification number of the plant unit. This information may be
used in the host for sorting alarms, etc.
2005 MODE_BLK
AUTO
The actual, target, permitted, and normal modes of the block
2006 BLOCK_ERR
-
This parameter reflects the error status associated with a block. It is
a bit-string, so that multiple errors can be shown
2007 UPDATE_EVT
-
The alert is generated by any change to the static data.
2008 BLOCK_ALM
-
The block alarm is used for all configuration error, hardware
connection failure or system problems in the block. The cause of
the alert is entered in the subcode field. The first alert to become
active will set Active status in Status attribute.
2009 TRANSDUCER_DIRECTORY
A directory that specifies the number and
starting indices of the transducers.
2010 TRANSDUCER_TYPE
Standard
conductivity
Conductivity transducer block
Conductivity
2011 XD_ERROR
-
The error code in transducer:
No failure,
Electronics failure,
I/O failure,
Mechanical failure
2012 COLLECTION_DIRECTORY -
A directory that specifies the number, starting indices and DD item Ids
of the data collection in each transducer within a transducer block
2013 PRIMARY_VALUE_TYPE
Conductivity conductivity /
Type of measurement represented by primary value
resistivity
2014 PRIMARY_VALUE
-
0 to 2 S/cm
Primary value of the instrument is Conductivity
2015 PRIMARY_VALUE_RANGE
0 to 2 S/cm 0 to 2 S/cm
The range of the instrument (can not change)
2016 SENSOR_CONST
0.1
0.005 to 50 cm
-1
The conductivity cell has a specific cell constant determined by the
dimensions of the cell
2017 CAL_POINT_HI
1.999
0 to 2 S/cm
Highest calibration point
2018 CAL_POINT_LO
0
0 to 2 S/cm
Lower calibration point
2019 CAL_MIN_SPAN
0,0001
> 0,0001 S/cm
Minimum span between two calibration points
2020 SENSOR_CAL_METHOD
-
1point, 2point
not used
2021 SENSOR_CAL_DATE
-
till 2104
Date the sensor was last calibrated
2022 SECONDARY_VALUE
-
-20 to 250 ºC ,
Temperature value
0 to 500ºF
2023 SECONDARY_VALUE_UNIT ºC
ºC, ºF
Temperature unit
2024 SENSOR_TEMP_COMP
automatic
automatic
2025 SENSOR_TEMP_MAN_VALUE -
-
No manual temp. value can be set. Always Automatic
2026 SENSOR_TYPE_TEMP
Pt1000
Pt1000, Pt100,
Temperature element used:
Ni100, Pb36, 8k55
2027 SENSOR_CONNECTION_
2
2
Only 2-wire connections supported
TEMP
2028 SENSOR_TYPE_COND
contact
2-electrode,
Either 2-electrode or 4-electrode contacting
2-electrode 4-electrode
conductivity cell can be selected
2029 SENSOR_OHMS
-
Actual cell resistance
2030 XD_MAN_ID
“ “
2031 TEMPERATURE_COEFF
2.1
0 to 3.5%/ºC (%/ºF) Process temperature compensation factor
2032 CONCENTRATION
-
-
Conductivity combined with temperature can be directly related to
the concentration. Concentraion is expressed in percentage
2033 TERTIARY_VALUE
-
0 to 2 S/cm
Second compensated conductivity value