Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual
Page 29

IM 12A00A01-61E
3-24 Foundation Fieldbus
The SIMULATE parameter of AI block consists of the elements listed in Table 3.23 below.
Table 3.23 SIMULATE Parameter
Sub-index
Parameters
Description
1
Simulate Status
Sets the data status to be simulated.
2
Simulate Value
Sets the value of the data to be simulated.
3
Transducer Status
Displays the data status from the transducer block. It cannot be changed.
4
Transducer Value
Displays the data value from the transducer block. It cannot be changed.
5
Simulate En/Disable
Controls the simulation function of this block.
1: Simulation disabled (standard)
2: Simulation started
When Simulate En/Disable in Table 3.23 above is set to 2, the applicable function block uses the simula-
tion value set in this parameter instead of the data from the transducer block. This setting can be used for
propagation of the status to the trailing blocks, generation of a process alarm, and as an operation test for
trailing blocks.