Page 98

IM 1E10C1-E
11-6
11. PARAMETER LIST
Indicator
Name
R/W
Description
Data Range, Units
BRAIN
Terminal
Decimal
Point
Default
Value
—
G00
4 - 20 SET
R
G01
G01
4 - 20 LOW CUT 0 to 10%
0
W
G02
G02
4 - 20 LOW LMT –20 to 100%
–20
W
G03
G03
4 - 20 H LMT
0 to 120%
120
W
—
G60
SELF CHECK
GOOD
R
to
EEPROM DEFAULT
—
H00
TEST
R
H01
H01
TEST MODE
(00) NORMAL
NORMAL
W
(01) TEST
H02
H02
OUTPUT VALUE –108 to 108
0
W
H03
H03
STATUS OUT
(00) NORMAL
NORMAL
W
(01) CLOSED (ON)
(02) OPEN (OFF)
—
H60
SELF CHECK
GOOD
R
to
EEPROM DEFAULT
—
L00
OTHER
R
L01
L01
TUNING
(00) INHIBIT
ENABLE
W
(01) ENABLE
L02
L02
KEY
00
0
W
55 << TO “N” >>
—
L60
SELF CHECK
GOOD
R
to
EEPROM DEFAULT
—
N00
APPL SET
n01
N01
TOTAL / PULSE (00) NO DAMP
DAMP
W
(01) DAMP
n02
N02
OUTPUT MODE (00) ON
ON
W
(01) OFF
n03
N03
RATE LIMIT
0 to 10%
5
W
n04
N04
DEAD TIME
0 to 15 s
0
W
n05
N05
POWER SYNCH (00) YES
YES
W
(01) NO
n07
N07
EMPTY PIPE
(00) ALARM
NO
W
(01) NO ALARM
ALARM
—
N60
SELF CHECK
GOOD
to
EEPROM DEFAULT
Current output functions.
Sets low cut width of current output.
Sets low limit of current output.
Sets high limit of current output.
Displays self-check result.
Output circuit test mode.
Selects test mode.
Sets test output value.
Selects test status output.
Displays self-check result.
Data setting restriction functions.
Inhibits or enables data setting with
indicator switches or BT200.
Enables “N” items display. Sets “55”
for access to “N” items.
Displays self-check result.
Special application.
Selects whether instantaneous flow rate
values or damped ones are used for
totalization and pulse output.
Selects active level of status output set
in “B10.” (except “ALARM OUT”)
Sets the level to reduce output fluctua-
tion.
Sets dead time to reduce output fluctua-
tion. When “0” is set, “Rate limit” is not
available.
Selects synchronization between excita-
tion frequency and power frequency.
Selects alarm effectiveness when flow
tube is not filled with fluid.
Displays self-check result.